西南石油大学学报(自然科学版) ›› 2006, Vol. 28 ›› Issue (2): 67-70.DOI: 10.3863/j.issn.1000-2634.2006.02.018

• 油气化学工程与化工 • Previous Articles     Next Articles

SPECTOPI30TOMETRIC FLOW-INJECTION ANALY-SIS ON TRACE I-IN RESERVOIR WATER

XIA0 Xin-feng ,ZHANG Xin-shen,LUO Ya-jun,et al .   

  1. ( Southwest Petroleum University,Cheng-du Sichuan 610065,China)
  • Received:2005-08-25 Revised:1900-01-01 Online:2006-04-20 Published:2006-04-20
  • Contact: XIA0 Xin-feng

Abstract: In the present method, a new , simple and rapid spectopho-
tometric flow injection technique for the accuracy and determina-
lion of I-in reservoir water has been developed. The method is
based on the decrease of absorbance intensity of rhodamine B
due to the complexation with[I2Br]-,I-can be selectively ox-
idated to form I2 by Ce(Ⅳ)in acidic medium and Iz reacts with
Br-to form[I2Br]-,At this moment, some changes of the so-
lotion's color takes place. It is found useful for sensitivity in CT-
MAB media. The absorption peak of the complexation, which is
increased linearly By addition of I-,occurs at 585 nm. Optimi-
zation of chemical and flow injection variables has been made.
Under the optimized conditions,the calibration curve obtained is
linear over the range 0. 0500~1. 00 mg/L, the linear equation
is H(mV) =76.6c(mg/L)-2. 933 3,r=0. 999 6,the de-
section limit is 0. 020 mg/L, and the precision(RSD=0. 85 %,
n=11 )is found quite satisfactory. This method has good sensi-
tivity and selectivity. Application of the method to the analysis
on I-in reservoir water resulted a good agreement between the
expected and found values.