J4 ›› 2016, Vol. 1 ›› Issue (8): 67-69.
Previous Articles Next Articles
Received:
Revised:
Online:
Published:
Abstract:
Objective: To understand the differences between whole blood(WB)mode and pre- dilution(PD)mode when using Sysmex XN1000, the automatic hematology analyzer, to test blood cells. Methods: To compare the precision of two modes in testing White blood cells(WBC), red blood cells(RBC), hemoglobin(Hb), hematocrit(HCT), platelets(PLT). Results: The precision of each test in whole blood mode corresponded with the requirements of precision in automatic blood analyzer;the precision of each testin predilution mode exceed the requirements of precision in automatic blood analyzer, t test showed that three group testing results of the two modes had significant differences in precision(P<0.05). The precision of PLT is the lowest. Conclusion: The whole blood mode, instead of the pre-dilution mode,should be adopted to test blood cells. The pre-dilution mode can only be adopted for infants and patients who are not suitable for venous blood collection.
Key words: Sysmex XN1000 hematology analyzer, whole blood mode, pre-dilution mode, clinical application
CLC Number:
R446.11
Song Zheng,Jin Zhuxuan. The Clinical Application of Testing Modes of Automatic Hematology Analyzer Sysmex XN1000[J]. J4, 2016, 1(8): 67-69.
0 / / Recommend
Add to citation manager EndNote|Ris|BibTeX
URL: http://journal15.magtechjournal.com/Jwk_dlxyzk/EN/
http://journal15.magtechjournal.com/Jwk_dlxyzk/EN/Y2016/V1/I8/67
〔1〕丛玉隆,王淑娟. 今日临床检验学〔M〕.北京:中国科学 技术出版社,1997:118. 〔2〕赵正军.迈瑞BC-5300血细胞分析仪两种操作模式间的 精密度分析〔J〕.实验与检验医学,2013,31(2):195-196. 〔3〕张宪华.血细胞分析仪自动全血和预稀释模式的结果分 析〔J〕.中国现代医生,2014,52(6):74-76. 〔4〕陈佩宣. 全自动血液分析仪全血模式与稀释模式的临床 应用〔J〕. 检验医学与临床,2012,9(12):1491-1492. 〔5〕罗来主,余建华,郭海燕.末梢血与静脉血分别在同一仪 器上作血细胞计数的探讨〔J〕.实验与检验医学,2011, 29(3):327 〔6〕钟金清,孟明,胡亚远.末梢血与静脉血在Abbott CD1800 血液分析仪上的结果比对分析〔J〕. 实验与检验医学, 2011,29(1):74. 〔7〕叶应妩,王毓三,申于瑜.全国临床检验操作规程〔M〕. 3版. 南京:东南大学出版社,2006:77-142. 〔8〕郑幼桥. MEK-6318K血液分析仪两种操作模式间精密 度分〔J〕.检验医学与临床,2007,4(11):1060-1061. 〔9〕唐友云,桂满元.冷凝集素对血常规的影响及不同处理 方法的分析〔J〕. 国际检验医学杂志,2014,35(17): 2401-2403. 〔10〕管晓媛,张丽,黄一玲,等. Coulter LH780 血细胞分析 仪自动和手动吸样模式检测结果比对分析〔J〕.中国卫 生检验杂志,2014,24(20):2889-2892.